81
Shandon Excelsior ES Operator Guide
The screen shows
The loading dates for concealed bottles and wax baths
•
The reagent usage count
•
The date of the last reagent rotation
•
The reagent usage count since the rotation
•
RISK OF CONTAMINATION OF PROCESSING
REAGENTS. USE COMMON SENSE WHEN
SELECTING WHICH REAGENT TO DRAW INTO THE
REACTION CHAMBER.
Highlight the reagent bottle.
Select
[iNSPECT REAGENT]
to move the reagent into the
reaction chamber.
You can
Sample the reagent
•
Assess for correct volume -
•
the reagent level is just above sensor 3 when the instrument is
loaded from 5L bottles. If using 3.8 L (1 US Gallon) bottles,
the chamber fluid level is just
below
sensor 3.
Add more reagent.
•
Select
[RETURN REAGENT]
to drain the chamber into the
same bottle. Or,
Inspecting a
Reagent