Shandon Excelsior ES Operator Guide
The following table describes items for regular checks.
AREA
ITEM
CHECK
WAX
BATHS
LEVEL
Check wax levels in all wax baths.
Carry out this check every day.
LID
SEAL
Always keep the lid and seal clear of solid wax deposits.
LATCH
Check that the lid latch operates correctly whenever
lid is opened.
The downdraft system should operate and the
reaction on-screen should show the lid open.
REAGENT
STORAGE
AREA
REAGENT
TUBES
When loading or unloading reagents, check that the
flexible parts of the reagent tubes in the front of the
cabinet are not damaged or collapsed.
REACTION
CHAMBER
HEATER TRIP
Press the red Heater Trip next to the Reset button in
the Reagent Storage Area. The Heater Reset switch
should operate and the reaction chamber Heater trip
icon will be displayed on the screen .
Press the Heater Reset switch fully in to reactivate
the reaction chamber heaters. The reaction chamber
Heater Fault icon will disappear from the screen.
Carry out this check every month.
BATTERY
ISOLATION
SWITCH
Check the operation of the switch. When the O side
is pressed (OFF), the battery fault icon should be
displayed. When the I side of the switch is pressed
(ON), the icon should disappear.
Carry out this check every month.
Regular
Maintenance
Содержание Shandon Excelsior ES
Страница 1: ...Shandon Excelsior ES Operator Guide English A78410120 Issue 3...
Страница 5: ...5 Shandon Excelsior ES Operator Guide...
Страница 12: ...12 Shandon Excelsior ES Operator Guide...
Страница 22: ...22 Shandon Excelsior ES Operator Guide...
Страница 62: ...62 Shandon Excelsior ES Operator Guide...
Страница 101: ...101 Shandon Excelsior ES Operator Guide...
Страница 103: ...103 Shandon Excelsior ES Operator Guide...
Страница 104: ...104 Shandon Excelsior ES Operator Guide...
Страница 112: ...112 Shandon Excelsior ES Operator Guide...
Страница 120: ...120 Shandon Excelsior ES Operator Guide Alcohols from dehydrants list Water Dilute detergent Water Cleaning Agents...