Product Manual for IonPac CS16
Page 27 of 65
5.5.
6-Cation Run (Li
+
, Na
+
, NH
4
+
, K
+
, Mg
2+
, Ca
2+
) at Room Temperature Operation
If the isocratic elution of ammonia, selected alkali metals, and selected alkaline earth cations (Li
+
, Na
+
, NH
4
+
, K
+
, Mg
2+
,
Ca2+) is run at room temperature using 30 mM Methanesulfonic acid (MSA), as shown below, potassium and magnesium
co-elute. By adjustment of the eluent concentration to 36 mM Methanesulfonic acid (MSA) separation of potassium and
magnesium is obtained. If better resolution between potassium and magnesium is required, the eluent should be increased by
2 mM MSA to 38 mM MSA. This higher eluent concentration will actually be more forgiving to room temperature
fluctuations, as shown in the following section.
Column:
IonPac CS16 5-mm Analytical Column (no guard)
Sample Volume:
25 µL Loop
Eluent: See
Chromatogram
Eluent Flow Rate:
1.0 mL/min
Temperature:
Room Temperature (approximately 22 °C)
Suppressor:
Cation Self-Regenerating Suppressor 300 (4-mm) in
AutoSuppression Recycle Mode
or MMS Suppressor:
Cation MicroMembrane Suppressor, CMMS 300 (4-mm)
MMS Regenerant:
TBAOH
MMS Mode:
Displacement Chemical Regeneration (DCR)
Expected Background Conductivity:
< 2 µS
Storage Solution:
Eluent
Eluent: 30 mM Methanesulfonic acid (MSA)
Analyte
mg/L
1. Lithium
0.1
2. Sodium
0.4
3. Ammonium
0.5
4. Magnesium
0.5
5. Potassium
1.0
6. Calcium
1.0
where 1 mg/L = 1 ppm
Eluent: 36 mM Methanesulfonic acid (MSA)
Figure 12
IonPac CS16 5-mm 6-Cation Run at Room Temperature
Document No. 031747-05
© 2010 Dionex Corporation
December 2010
Содержание IonPac CG16
Страница 2: ...IonPac CS16 IonPac CG16 for ...
Страница 62: ...Product Manual for IonPac CS16 Page 60 of 65 Document No 031747 05 2010 Dionex Corporation December 2010 ...
Страница 63: ...Product Manual for IonPac CS16 Page 61 of 65 Document No 031747 05 2010 Dionex Corporation December 2010 ...
Страница 64: ...Product Manual for IonPac CS16 Page 62 of 65 Document No 031747 05 2010 Dionex Corporation December 2010 ...