Page 21 of 65
SECTION 5 – EXAMPLE APPLICATIONS
The chromatograms in this section were obtained using columns that reproduced the Production Test Chromatogram (see
Section 5.3, “Production Test Chromatogram”) on optimized Ion Chromatographs (see Section 3, “Installation”). Different
systems will differ slightly in performance due to slight differences in column sets, system void volumes, liquid sweep-out
times of different components and laboratory temperatures.
Before attempting any of the following example applications, take the time to ensure that your system is properly configured.
Ensure that all of the eluents have been made from high purity reagents and deionized water. All water used in the
preparation of eluents should be degassed, deionized water. For chemical purity requirements, see Section 4.3, “Chemical
Purity Requirements.” After running synthetic standards to calibrate your system, you may find that real sample matrices foul
your columns. For this reason it is always advisable to use a guard column to protect the analytical column. If column
performance deteriorates and it is determined that the guard or the analytical column has been fouled, refer to the column
cleanup protocols in, “Column Care.” If your sample matrices are relatively low in ionic concentration, you may be able to
increase the sensitivity of your system by using sample concentration techniques (see Section 4.5, “Sample Concentration”).
Document No. 031747-05
© 2010 Dionex Corporation
December 2010
Содержание IonPac CG16
Страница 2: ...IonPac CS16 IonPac CG16 for ...
Страница 62: ...Product Manual for IonPac CS16 Page 60 of 65 Document No 031747 05 2010 Dionex Corporation December 2010 ...
Страница 63: ...Product Manual for IonPac CS16 Page 61 of 65 Document No 031747 05 2010 Dionex Corporation December 2010 ...
Страница 64: ...Product Manual for IonPac CS16 Page 62 of 65 Document No 031747 05 2010 Dionex Corporation December 2010 ...