Product Manual for IonPac CS16
Page 26 of 65
5.4.
6-Cation Fast Run (Li
+
, Na
+
, NH
4
+
, K
+
, Mg
2+
, Ca
2+
) at 40 °C
Isocratic elution of ammonia, selected alkali metals, and selected alkaline earth cations. (Li
+
, Na
+
, NH
4
+
, K
+
, Mg
2+
, Ca
2+
) can
be completed within 10 minutes. Notice that with this higher eluent concentration, magnesium, a divalent, elutes before
potassium, a monovalent. This is an ideal operation mode when analytes are present in similar concentrations.
Column:
IonPac CS16 5-mm Analytical Column (no guard)
Sample Volume:
25 µL Loop
Eluent:
48 mM Methanesulfonic acid (MSA)
Eluent Flow Rate:
1.0 mL/min
Temperature: 40
°C
SRS Suppressor:
Cation Self-Regenerating Suppressor 300 (4-mm) in
AutoSuppression Recycle Mode
or MMS Suppressor:
Cation MicroMembrane Suppressor, CMMS 300 (4-mm)
MMS Regenerant:
TBAOH
MMS Mode:
Displacement Chemical Regeneration (DCR)
Expected Background Conductivity:
< 2 µS
Storage Solution:
Eluent
Analyte
mg/L
1. Lithium
0.1
2. Sodium
0.4
3. Ammonium
0.5
4. Magnesium
0.5
5. Potassium
1.0
6. Calcium
1.0
where 1 mg/L = 1 ppm
0
5
10
0
0.2
0.4
0.6
0.8
1.0
1.2
1.4
1.6
1.8
2.0
5
4
6
1
3
2
0
µS
1.8
10
0
Minutes
5
Figure 11
IonPac CS16 5-mm 6-Cation Fast Run at 40 °C
Document No. 031747-05
© 2010 Dionex Corporation
December 2010
Содержание IonPac CG16
Страница 2: ...IonPac CS16 IonPac CG16 for ...
Страница 62: ...Product Manual for IonPac CS16 Page 60 of 65 Document No 031747 05 2010 Dionex Corporation December 2010 ...
Страница 63: ...Product Manual for IonPac CS16 Page 61 of 65 Document No 031747 05 2010 Dionex Corporation December 2010 ...
Страница 64: ...Product Manual for IonPac CS16 Page 62 of 65 Document No 031747 05 2010 Dionex Corporation December 2010 ...