Product Manual for IonPac CS16
Page 17 of 65
4.3.2.
Inorganic Chemicals
Reagent Grade inorganic chemicals should always be used to prepare ionic eluents. Whenever possible, inorganic chemicals
that meet or surpass the latest American Chemical Society standard for purity should be used. These inorganic chemicals will
detail the purity by having an actual lot analysis on each label. The following chemicals will perform reliably:
A.
Use Fluka or Aldrich Methanesulfonic Acid (MSA) (>99% pure) or Dionex Methanesulfonic Acid (0.4 M).
B.
Use only concentrated sulfuric acid (H
2
SO
4
), ACS grade or BAKER INSTRA-ANALYZED
®
for trace metals.
C.
Use Dionex Cation Regenerant Solution, tetrabutylammonium hydroxide (TBAOH), P/N 039602, to ensure
maximum system performance when operating with a CMMS 300, or a CSRS 300 in the Chemical Suppression
Mode. For the DCR Mode, use Dionex TBAOH (P/N 057561).
D.
Use deionized water with a specific resistance of at least 18.2 megohm-cm to make all standards, eluents and
regenerants.
4.4.
Preparation of Eluent Stock Solution Concentrates
Sulfuric acid (H
2
SO
4
) is very corrosive. Methanesulfonic acid (MSA) is also a corrosive and a strong
irritant.
WARNING
!
Avoid breathing the vapors.
Always use these reagents in a fume hood. Wear gloves and goggles.
4.4.1.
1.0 N Methanesulfonic Acid (MSA) Stock Solution
A 1.0 N methanesulfonic acid stock solution can be prepared as follows:
A.
Weigh out 96.10 g of methanesulfonic acid (MSA, > 99%, P/N 033478).
B.
Carefully add this amount to a 1-liter volumetric flask containing about 500 mL of deionized water.
C.
Dilute to the mark and mix thoroughly.
4.4.2
0.4 N Methanesulfonic Acid (MSA) Eluent Concentrate
0.4 N Methanesulfonic Acid Eluent Concentrate (P/N 057562 or package of 4, P/N 057568) is available from Dionex
4.4.2.
1.0 N Sulfuric Acid Stock Solution
This solution will be used in the preparation of each of the eluents in Section 5, “Example Applications.”
Calculate the amount (in grams) of concentrated sulfuric acid (H
2
SO
4
) that you need to add to a 1 liter volumetric flask by
using the % H
2
SO
4
composition stated on the label of the particular bottle of H
2
SO
4
you are using. For example, if the H
2
SO
4
concentration is 98%, you need to weigh out 50.04 grams of concentrated H
2
SO
4
. Carefully add this amount of H
2
SO
4
to a 1-
liter volumetric flask containing about 500 mL of deionized water with a specific resistance of 18.2 megohm-cm. Dilute to
the 1 liter mark and mix thoroughly.
In other words:
FW
of
H
2
SO
4
= 98.08 g
H
2
SO
4
concentration = 98%
Document No. 031747-05
© 2010 Dionex Corporation
December 2010
Содержание IonPac CG16
Страница 2: ...IonPac CS16 IonPac CG16 for ...
Страница 62: ...Product Manual for IonPac CS16 Page 60 of 65 Document No 031747 05 2010 Dionex Corporation December 2010 ...
Страница 63: ...Product Manual for IonPac CS16 Page 61 of 65 Document No 031747 05 2010 Dionex Corporation December 2010 ...
Страница 64: ...Product Manual for IonPac CS16 Page 62 of 65 Document No 031747 05 2010 Dionex Corporation December 2010 ...