Product Manual for IonPac CS16
Page 22 of 65
5.1.
Production Test Chromatogram at 40 °C
Isocratic elution of cations on the IonPac CS16 Analytical Column has been optimized utilizing a methanesulfonic acid
eluent. Using this eluent, the weak carboxylate functionalized packing isocratically separates mono- and divalent cations in a
single injection. To guarantee that all IonPac CS16 Analytical Columns meet high quality and reproducible performance
specification standards, all columns undergo the following production control test.
Column: CS16
5-mm
Sample Volume:
25 µL Loop
Eluent:
30 mM Methanesulfonic acid (MSA)
Eluent Flow Rate:
1.0 mL/min
Temperature: 40
°C
Suppressor:
Cation Self-Regenerating Suppressor 300 (4-mm)
in AutoSuppression Recycle Mode,
or MMS Suppressor:
Cation MicroMembrane Suppressor, CMMS 300 (4-mm)
MMS Regenerant:
TBAOH
MMS Mode:
Displacement Chemical Regeneration (DCR)
Expected Background Conductivity:
< 2 µS
Storage Solution:
Eluent
0
5
10
15
20
25
0
0.2
0.4
0.6
0.8
1.0
1.2
1.4
1.6
1.8
5
4
6
1
3
2
0
µS
1.6
10
0
15
Minutes
5
20
25
Analyte
mg/L
1. Lithium
0.1
2. Sodium
0.4
3. Ammonium
0.5
4. Potassium
1.0
5. Magnesium
0.5
6. Calcium
1.0
where 1 mg/L = 1 ppm
IonPac CS16 5-mm
Analytical Column Only
0
5
10
15
20
25
0
0.2
0.4
0.6
0.8
1.0
1.2
1.4
1.6
1.8
2.0
5
4
6
1
3
2
0
µS
1.6
10
0
15
Minutes
5
20
25
IonPac CS16 & CG16 5-mm
Analytical & Guard Columns
Figure 7
IonPac CS16 5-mm Production Test Chromatogram
Document No. 031747-05
© 2010 Dionex Corporation
December 2010
Содержание IonPac CG16
Страница 2: ...IonPac CS16 IonPac CG16 for ...
Страница 62: ...Product Manual for IonPac CS16 Page 60 of 65 Document No 031747 05 2010 Dionex Corporation December 2010 ...
Страница 63: ...Product Manual for IonPac CS16 Page 61 of 65 Document No 031747 05 2010 Dionex Corporation December 2010 ...
Страница 64: ...Product Manual for IonPac CS16 Page 62 of 65 Document No 031747 05 2010 Dionex Corporation December 2010 ...