Product Manual for IonPac CS16
Page 25 of 65
5.3.
6-Cations (Li
+
, Na
+
, NH4
+
, K
+
, Mg
2+
, Ca
2+
) with CS16 3-mm and CAES
In order to use a CS16 3-mm column with an Atlas Suppressor, the maximum eluent concentration allowed is 25 mM MSA.
Column:
IonPac CS16 3-mm Analytical Column (no guard)
Sample Volume:
25 µL Loop
Eluent:
25 mM Methanesulfonic acid (MSA)
Eluent Flow Rate:
0.36 mL/min
Temperature: 40
°C
AES Suppressor:
Cation Atlas Electrolytic Suppressor, CAES
or SRS Suppressor:
Cation Self-Regenerating Suppressor, CSRS 300 (2-mm)
or MMS Suppressor:
Cation MicroMembrane Suppressor, CMMS 300 (2-mm)
MMS Regenerant:
TBAOH
MMS Mode:
Displacement Chemical Regeneration (DCR)
Expected Background Conductivity:
< 0.5 µS
Storage Solution:
Eluent
0
5
10
15
20
25
30
35
Min u tes
0.10
1.90
µS
3
1
2
4
5
6
Analyte mg/L
1. Lithium
0.05
2. Sodium
0.20
3. Ammonium
0.25
4. Potassium
0.50
5. Magnesium
0.25
6. Calcium
0.50
where 1 mg/L = 1 ppm
Figure 10
IonPac CS16 3-mm 6-Cation Run at 40 °C with an Atlas Suppressor
The Cation Atlas Electrolytic Suppressor (P/N 056118) can be used with up to 25 mM Methanesulfonic acid at 1.0 mL/min,
as with a CS16 5-mm column.
Document No. 031747-05
© 2010 Dionex Corporation
December 2010
Содержание IonPac CG16
Страница 2: ...IonPac CS16 IonPac CG16 for ...
Страница 62: ...Product Manual for IonPac CS16 Page 60 of 65 Document No 031747 05 2010 Dionex Corporation December 2010 ...
Страница 63: ...Product Manual for IonPac CS16 Page 61 of 65 Document No 031747 05 2010 Dionex Corporation December 2010 ...
Страница 64: ...Product Manual for IonPac CS16 Page 62 of 65 Document No 031747 05 2010 Dionex Corporation December 2010 ...