C. Use the following gradient program to remove hydrophobic cations and organic contaminants.
Eluent 1:
100 mM HCl
Eluent 2:
90% Acetonitrile in deionized water
Time (min)
% E1
% E2
0.0
100
0
20.0
0
100
25.0
0
100
45.0
100
0
55.0
100
0
D.
Rinse the column for 15 minutes with 10 mM HCl
before pumping eluent over the column.
E.
Equilibrate the column(s) with eluent
for at least 30 minutes before resuming normal operation.
F.
Reconnect the IonPac CS16.
Connect the Analytical/Capillary Column outlet to the Suppressor and the inlet to
either the IonPac CG16 Guard Column or the Pump Module.
G. Equilibrate the column with eluent before resuming normal operation.