77210163GB Issue 3
16
2.4
MESSAGE CODES
The following displays show in the Temperature display when conditions are
appropriate.
CHAMBER CONTROLS
CRYOBAR
BOOST
WINDOW
LOCK
LAMP
CRYOBAR
TEMP
WORKING
TEMP
IMMED.
DEFROST
IMMED.
FUMIGATE
CANCEL
DEFROST/
FUMIGATE
TEMPERATURE °C
+ -
i
Ald shows when fumigation is in progress.
ii
LOBAT signifies that the internal back-up battery is not charged sufficiently to
retain information during an interruption of the mains supply.
The battery retains programmed information in memory while the power is switched
off. Stored information is lost if it is allowed to discharge.
Power must be on to recharge the battery. If
LOBAT still shows after power is
restored for more than 12 hours, contact your Thermo dealer for service.
iii
dEf shows when defrosting is in progress.
iv
--- shows if the associated temperature probe is faulty.
v Fault Codes show in the Temperature display if the instrument malfunctions. Press
[wORKING TEMP] or [CRYOBAR TEMP] to over-ride most faults. Faults F:11,
F:12 and F:13, cause the instrument to wait for 30 mins then reset and attempt to
reach its default temperature of -20°C.
Note
1 If the fault fails to clear, contact your Thermo Dealer for service.