77210163GB Issue 3
24
Slide the handwheel carefully onto the drive shaft at the top
of the right hand side panel. The handwheel is supplied in
the Accessory Pack. Make sure that the locating pin of the
handwheel fits into the key-way of the drive shaft.
The lever next to the handwheel handle is used to lock
the handwheel to prevent inadvertent rotation. To lock the
handwheel, push the lever away from you.
Fit the handwheel bolt then use the special Handwheel Bolt
Key to turn the retaining bolt clockwise to tighten. Rotate the
handwheel and check that movement is smooth and produces
a corresponding movement of the Specimen Head.
Turn the handwheel until the handle is at the 12 o’clock position
then push the lever of the lock away from you. Check that the
lock operates and that the handwheel cannot rotate. Check that
the Specimen Head is at the top of its travel.
Move the lock lever towards you and check that the handwheel
rotates smoothly and produces vertical oscillations of the
Specimen Head. Check also that the lock operates at the ‘quarter
past’ and ‘half past’ the hour clock positions.
Slide the waste bottle retaining bar to the right, and remove
the Waste Bottle. Use a 10% solution of Formalin to fill the
Waste Bottle sufficient to cover the outlet tube - approximately
15mm (½ inch) deep. The lid of the waste bottle turns counter
clockwise to undo.