77210163GB Issue 3
22
j
k
Microtome Transit fixings: [E only]
From the underside, support the stud of the clamping
bracket that prevents damage to the microtome
leadscrew, then undo and remove the top nut (h).
Support the stud from underneath while allowing it to fall
clear of the leadscrew arm.
Support the leadscrew arm and use the nylon loop
(i) to remove the spacer from between the arm of the
leadscrew and the casting of the microtome.
Check that the detection flag at the end of the arm is in
line with the slots in the two black plastic opto units of the
printed circuit board (j). This signifies that the leadscrew
and trunnion nut are located correctly.
Note
1 Turn the leadscrew to adjust the setting if
necessary.
Undo and remove the transit plate that supported the
defrost valve (k). Replace the rear panel and 8 screws.
Retain the transit plate for future use when moving the
Cryotome.
3.4 TO INSTALL ThE CRYOTOME
Move the instrument to its permanent location. This must
be
level and provide a 300 mm (12”) clearance to the
sides and rear of the instrument for ventilation.
300mm