77210163GB Issue 3
44
ALwAYS wIPE UP ANY SPILLS IMMEDIATELY. IN ThE EVENT Of A
MAJOR SPILLAGE, DISCONNECT ThE INSTRUMENT fROM ThE MAINS
SUPPLY wIThOUT DELAY AND DO NOT RECONNECT AND SwITCh ON
UNTIL ThE INSTRUMENT hAS BEEN ThOROUGhLY DRIED OUT AND
ChECKED BY A SERVICE ENGINEER.
POTENTIALLY LEThAL VOLTAGES ABOVE 110V a.c. ARE PRESENT
INSIDE ThE UNIT. DO NOT REMOVE ANY ACCESS COVERS
Always wear protective gloves when you clean or decontaminate the
Cryotome to protect yourself against the effects of chemicals
Do not use any chemicals that may interact with materials of
manufacture. If in doubt, check with Thermo Service department
Do not use abrasive compounds or metal components to clean the
Cryotome or its components and accessories
5.2
fUMIGATION
Carefully load the heating trough in the Refrigerated
Chamber with 2 ml of concentrated formalin solution. Close
the window .
Press
[wINDOw LOCK] and check that the push-button
indicator is lit.
To start the fumigation immediately, or to program in a
start time, press
[IMMED fUMIGATE] in the Chamber
Controls panel.
Fumigation takes place at the same time every day if the ‘DAY 8’ option is selected in
the Fumigation program. Before the scheduled time for fumigation each day it is
therefore essential that:
i
2 ml of concentrated formalin is in the heater trough;
wARNING
wARNING
wARNING
CHAMBER CONTROLS
CRYOBAR
BOOST
WINDOW
LOCK
LAMP
CRYOBAR
TEMP
WORKING
TEMP
IMMED.
DEFROST
IMMED.
FUMIGATE
CANCEL
DEFROST/
FUMIGATE
TEMPERATURE °C
+ -
CHAMBER CONTROLS
CRYOBAR
BOOST
WINDOW
LOCK
LAMP
CRYOBAR
TEMP
WORKING
TEMP
IMMED.
DEFROST
IMMED.
FUMIGATE
CANCEL
DEFROST/
FUMIGATE
TEMPERATURE °C
+ -
ALd