77210163GB Issue 3
13
vii the instrument reverts to its previous setting when the fumigate cycle is
finished.
Note
1 If the Day 9 option is selected when the fumigation time is set, then the instrument
enters Standby (PSAVE) mode when fumigation finishes.
2.2.4.3 [CANCEL DEfROST/fUMIGATE]
Press
[CANCEL DEfROST/fUMIGATE] to terminate a Defrost or Fumigate sequence.
The button LED remains lit until the process has finished.
2.2.4.4 [LAMP]
Press
[LAMP] to switch on the fluorescent lamp to illuminate the work area. The button
LED is lit when the light is on. The next press switches off the lamp and indicator.
2.2.4.5 [CRYOBAR BOOST]
Press
[CRYOBAR BOOST] to turn on the peltier element in the Cryobar for 10 minutes.
The compressor is also switched on for this period. (However, if the compressor
switched off within the last three minutes, there is a delay of up to three minutes before
the compressor re-starts.)
If
[CRYOBAR BOOST] is pressed while its LED is flashing at the end of the boost
period, the boost is repeated for a further 10 minutes, otherwise normal operation
follows when a boost period times out.
Note
1 The Cryobar is situated on the refrigeration pipework and its temperature can
vary considerably. Do not use the Cryobar for storing specimens.
2.2.4.6 [wINDOw LOCK]
[wINDOw LOCK] locks the window in the closed position. The window must be fully
closed before the lock can operate. A second press releases the lock. If the window is
not properly closed when the button is pressed,
Err 4 is displayed. The lock remains in
the locked state during power off (if already locked).