77210163GB Issue 3
10
2.2.3.1 TO SET ThE CLOCK
Press and hold
[CLOCK] and then press the relevant button detailed below. Release the
buttons when the correct information is displayed.
To Set the...
Press and hold
[CLOCK] then press:
Display will show:
Day
[DAY]
1 - 7
Note Keep Monday as
Day 1 to avoid confusion
Hours
[hOURS]
0 - 23
Minutes
[MINUTES]
1 - 59
2.2.3.2 TO SET DEfROST TIMES
Press and hold
[DEfROST] and then press the relevant button detailed below. Release
the buttons when the correct information is displayed.
To Set the...
Press and hold
[DEfROST] then
press:
Display will show:
Day
[DAY]
0 - 9 (see note 1 below)
Hours
[hOURS]
0 - 23
Minutes
[MINUTES]
1 - 59
Note:
1 Use the following settings for the days:
DAY 0
= prevents the programmed function.
DAY 1-7
= defrosting starts at the time set on the day shown (set
Monday as Day 1 to avoid confusion).
DAY 8
= defrosting starts at the set time every day
DAY 9
= STANDBY (PSAVE = power save)
2 Standby (PSAVE) allows the instrument to remain switched on in a defrosted
state for up to 24 hours until a pre-determined time. It then returns to its normal
operational state
To set Standby (PSAVE) mode, press and hold [DEfROST] then press [DAY] and
select day 9. Set the time that normal operation is required to resume. Release the
buttons.
Содержание Cryotome
Страница 72: ...77210163GB Issue 3 72 ...