77210163GB Issue 3
47
Note
1 It is not necessary to use formalin during the Defrost/Fumigate cycle
The waste bottle collects water from the cooling coils of the heat exchanger.
It also collects debris when the Refrigerated Chamber is washed down so the
contents must be disposed of in an appropriate manner.
ALwAYS KEEP AT LEAST 200ml Of 10% fORMALIN SOLUTION OR
SIMILAR DE-CONTAMINANT IN ThE wASTE BOTTLE. MAKE SURE
ThAT ThE END Of ThE DRAIN TUBE IS KEPT SUBMERGED.
If condensation appears on the window in conditions of high
relative humidity, set the Window Boost switch at the rear of the
instrument to ‘ON’ .
Note
1 The window boost facility provides extra de-misting power in conditions of high
humidity
2 The refrigeration system operates less efficiently if the window boost facility is
left switched on unnecessarily
5.5
MAINTENANCE
SwITCh Off ThE INSTRUMENT AT ThE MAINS AND REMOVE ThE
PLUG fROM ThE wALL SOCKET BEfORE PROCEEDING wITh ANY
MAINTENANCE.
5.5.1
TO REMOVE ThE LAMP TUBE
Lift up the top cover.
Loosen the two screws that hold the shade.
Slide the shade along its two keyhole slots, then lift it
off.