[E only]: Alternatively, if sufficient ‘Travel Available’ is indicated on the Specimen Travel
Indicator, advance the specimen towards the knife using the double chevron (coarse),
then the single chevron (fine), push-buttons of the miniature control panel on the left
hand side.
If necessary use the Rotary Control to obtain fine positioning of the specimen with
respect to the knife. The Rotary Control moves the specimen 8 microns (µm) with each
‘click’.
[Standard only]: Advance the specimen towards the knife using the Rewind/Advance
knob on the Advance Control Panel.
4.4 TO CUT SECTIONS
KEEP ThE wINDOw CLOSED whEN CUTTING OR INJURIOUS AEROSOLS
MAY BE INhALED.
At the Micron Selector panel use the
[+] and [–] push-buttons [E only] or the Micron
Selector Knob [Standard only] to set the required section thickness.
MICRON SELECTOR
+ -
[E only]
Std only]
To cut sections, turn the Handwheel smoothly. The instrument performs a single cutting
stroke with each full revolution of the Handwheel and cuts a section of the specimen.
Continue to turn the Handwheel until the required number of sections have been cut.
The Specimen automatically advances by the set distance before the start of each
cutting stroke.