TROUBLE ShOOTING
6.1 GENERAL
Correct service and maintenance is essential for the long term serviceability of
precision engineered products such as the Cryotome. We strongly recommend that
a Thermo Service Contract is used to ensure future reliability, and consistency of
performance.
Cryotome has self test routines built in so that fault codes
(f:(Number)) show in the
displays if malfunctions occur. No operator response is involved following a fault code
other than to contact your Thermo Engineer for service.
Table 1 lists basic instrument function problems and solutions.
Table 2 lists other problems and their causes that are not shown in the displays.
Table 3 lists alarm signals, their causes, and the remedial action to be taken.
If a fault indication continues or recurs, remove the samples from the refrigerated
chamber and store them in a suitable refrigerated container and contact your Thermo
Service Representative.
TABLE 1
INSTRUMENT fUNCTION
SYMPTOM
CAUSE
REMEDY
The Cryotome does
not respond when
the mains power is
switched on
1 The instrument is still
carrying out initial tests
2 No mains supply
3 The mains fuses have
blown
1 Wait 30 seconds for the
initial tests to finish
2 Connect the power lead
Switch on the mains power
and instrument main power
switch
(see paragraphs 3.4 and 3.5)
3 Replace the mains fuse
Replace the instrument
fuses
(Note only a technically
competent person should
replace fuses)
The
Cryotome
is
powered up, but does
not respond correctly
1 The Cryotome may
have started incorrectly
1 Switch off the instrument.
Wait for 30 seconds. Switch
on the main power switch.