77210163GB Issue 3
35
Ensure that there is a good edge i.e. there are no ‘nicks’
- on the anti roll plate. Use the brush supplied to keep the
anti roll plate, the blade, and the clamp clean.
ThE EDGES Of ThE GLASS ANTI-ROLL
PLATE ARE ShARP
Note
1 To prevent the anti-roll plate heating up, make
sure that it is kept close to the knife
3.6.6
SPECIMEN hEAD
The Specimen Head is designed to accept a Thermo
Cryocassette that already has a frozen specimen mounted.
The frozen specimen on the Cryocassette is held in the
Specimen Head by a Quick Release Cryocassette Clamp.
The Clamp lever (b) at the top of the Specimen Head
operates a moveable jaw that holds the Cryocassette
firmly in position.
To install a cryocassette in the Quick Release Clamp:
i
Pull the Clamp lever (b) towards you.
ii
Fit the cryocassette under the two fixed jaws as
shown.
iii Release the Clamp lever.
3.6.6.1 fixed Orientation head (supplied with instrument)
To secure the cryocassette in the Fixed Head, insert the
shaft on the Quick Release Clamp into the hole in the Fixed
Head and tighten the knob on top of the Fixed Head (c).