49
4. Carefully and cleanly cut the EASYpure R
O
D
I
feed-
water tubing such that one piece will connect the
EASYpure R
O
D
I
to the external accessory pump.
The other piece will be used to connect the facility
water supply to the “inlet” of the accessory pump.
Use 90-degree fittings as necessary to help route
tubing to and from the external pump accessory.
5. Attach power cable from back of EASYpure R
O
D
I
into the side of the Pump Housing.
6. Open facility water supply valve and check
AY1332X1 for leaks.
7. Turn power on to the EASYpure R
O
D
I
system and
operate normally.
AY1332X2 External Filter
Accessory
This accessory should be used when customer site has feed-
water with high particulate and/or high scaling contaminants.
If in doubt contact Technical Service for possible feedwater
testing.
Accessory Kit Includes:
1. Filter housing (1)
2. Plumbing adapters (2)
3. Pressure gauges (2)
4. Filter/Antiscalant assembly (1)
5. Wall bracket (1)
6. Screw/washer sets for bracket/housing connection (4)
Installation Instructions
1. Turn power off to the EASYpure R
O
D
I
system.
2. Close facility water supply valve.
A