Troubleshooting
Problem
Possible Causes
Action
EASYpure
R
O
D
I
completely inactive.
No electrical power to
Ensure that the EASYpure
R
O
D
I
(Pump not operating, display not lit, etc.)
EASYpure
R
O
D
I
.
power cord is connected to a live
power source and completely plugged
into electrical outlet.
Note: When unit is powered and not
Membrane key switch leads not
Disconnect unit from power.
operating, nor in standby, “IdL”
connected.
Check and reconnect.
will normally be displayed.
Main power switch off.
Place to “On” position.
Fuses blown or not installed
Check to make sure proper fuses
properly.
were installed.
Replace the fuses as indicated in
the
Fuse Replacement
section.
Pump does not run. Display showing
Loose wire connection to pump.
Ensure pump is properly connected
purity information.
to display/control (PC1286X3).
Display/control board is defective.
Call Technical Service.
Pump worn out or defective.
Replace pump. Call Technical
Service.
Recirculated water will not rinse
Exhausted cartridge
Replace the cartridge as
up to desired purity level.
indicated in the
Cartridge
Replacement
section.
Cartridges out of order.
Install the cartridges in the
proper order as indicated in
the
Cartridge Installation
section.
Dirty resistivity cell.
Clean resistivity cell as indicated in the
Maintenance and Servicing
section.
Water path restriction.
Check tubing for constriction
(kinks) or blockages.
Resistivity meter board out of
Replace board (PCX70). Call
tolerance.
Technical Service.