Atmospheric Drain
The RO reject and flush water is sent to drain through this
connection.
1. Locate the drain water tubing. This is the 1/4”
O.D. tubing that is approximately 6 ft. long with
a 1/4” O.D. x 1/4” N.P.T. tubing adapter on one
end. The atmospheric drain fitting is located on
the lower left corner in the rear of the EASYpure
R
O
D
I
.
2. Install the tubing adapter into the atmospherical-
ly vented drain and route the tubing to the
EASYpure R
O
D
I
drain connector, ensuring that
there are no kinks
3. Take the tubing end that has no fitting, wet the
tube end with water and insert the tube straight
into the unit drain connector until it bottoms out.
Refer to section
Push-to-Connect Fitting
Tubing Installation
for more explicit details.
Overflow Drain
1. To extend tank overflow tubing to an atmospher-
ic drain, use 1/2" I.D. tubing and tubing connec-
tors (user supplied) to connect the overflow
drain tubing (lower left corner of the rear of the
unit) to an atmospherically vented sink or floor
drain.
19
I
NSTALLATION
Figure 11: EASYpure R
O
D
I
Back
(External Unit Connections)
Door Latch
P
o
we
r to
Ex
t
e
r
n
a
l B
oo
s
t
Pump A
cc
ess
or
y
R
ese
r
v
o
i
r Dra
i
n Plu
g
I
nle
t
W
at
e
r
C
o
nne
ct
i
o
n
A
t
m
o
sp
h
e
r
i
c Dra
i
n
O
v
e
r
f
l
o
w
START/STOP
M
-cm
RO
S
yste
m
O
perating
Low
RO
P
u
rity
E
AS
Y
p
u
re
R
o
DI
D
isp
l
ay
F
i
ll
P
o
rt
V
entgar
d C
ap
A
ir
b
o
rne
F
i
l
ter
D
ra
w
-
o
ff
V
a
l
v
e
L
e
v
er
S
ig
h
t
G
a
u
ge
Figure 10: EASYpure R
O
D
I
Front