3. Completely and securely assemble AY1332X2 com-
ponents. See drawing on page 53. (Note orientation
of Filter/Antiscalant assembly). The use of
PTFE
tape on NPT treads is recommended to aid in a
secure leak proof assembly.
4. Using 2 customer-supplied fasteners, install the
assembly on a wall in a location convenient for filter
exchange. Be sure NOT to mount the assembly
over anything that could be damaged or be a safety
issue if water spillage occurs.
5. Install the D13321 inlet tubing connection into the
outlet plumbing adapter.
6. Using a customer-supplied tubing/fitting (1/4 inch
male NPT), securely attach feedwater connection to
the 1/4 inch female NPT inlet plumbing adapter.
7. Open facility water supply valve and check
AY1332X2 assembly for leaks.
8. Turn power on to EASYpure R
O
D
I
system and oper-
ate normally.
Filter/Antiscalant Cartridge Exchange
Filter/Antiscalant cartridge should be exchanged whenever
inlet-outlet pressure difference exceeds 10psi (0.69 bar).
1. Turn power off to the EASYpure R
O
D
I
system.
2. Close/shut off facility water supply or supply valve
(located upstream of the AY1332X2 assembly.)
3. Depressurize housing by momentarily pressing the
RED button on top of the housing cover. Note that a
very small amount of water may spray out of the
depressurization valve, under the RED button, dur-
ing this step.
4. Carefully unscrew housing from housing cover, as it
will be filled with water.
5. Remove and discard old Filter/Antiscalant cartridge.
A
CCESSORIES