10. Close the cartridge access door.
11. Install the hose barb into the fitting in the draw-
off valve.
12. Turn unit on and from “IdL” press the
START/STOP button to begin unit operation.
13. Rinse approximately 3 liters (1/2 reservoir vol-
ume) of water through the draw-off valve to
drain.
14. Close the draw-off valve.
15. Remove the hose barb from the draw-off valve
and proceed to the
0.2 Micron Final Filter
Replacement
section of this manual.
16. Place unit in Run Mode until desired purity is
achieved.
UV Lamp Replacement (See Fig. 5)
Lamp life will vary according to the number of times the
EASYpure R
O
D
I
unit is turned on and off. The UV lamp
should be replaced a minimum of every 12 months of
operation. If the EASYpure R
O
D
I
is cycled between Run,
Standby and Idle Modes frequently during the work day,
this may result in a shorter lamp life. Therefore, it is rec-
ommended that the EASYpure R
O
D
I
be left in the Run
Mode during normal working hours and in Standby Mode
at night and on weekends.
To replace the ultraviolet lamp:
1. Turn off the EASYpure R
O
D
I
and disconnect it
from the power supply. Remove the power
cord.
2. Depressurize the system by opening draw-off
valve and allowing water to drain from the unit.
Disconnect water line.
3. Remove the reservoir Ventgard cap.
M
AINTENANCE AND
S
ERVICING
Warning
This unit is equipped with an ultraviolet
lamp. Ultraviolet radiation is harmful to
the eyes and skin. Do not attempt to
observe the lamp directly, while it is
illuminated.
Caution
Do not unscrew the metal end cap
cover of the UV chamber, as this will
loosen the water tight seal and may
damage the replacement lamp. Pull the
black plastic cover straight up.
Warning
Depressurize system prior to removing
cover.
35