Storage Reservoir Filling and
Cartridge/Membrane Rinse Up
RO Membrane Rinse-Up:
1. During initial operation be sure the reservoir
Drain Plug (bottom rear of unit) is removed and
that the unit is placed such that the reservoir
drain is over a sink to allow for proper rinsing
of RO membrane.
2. Connect water supply to unit by inserting the
quick disconnect into the feed water inlet on
the lower left side in the rear of the EASYpure
R
O
D
I
. Turn on main water supply. Turn the
system power on by depressing the Main
Power Switch to the ʻIʼ position. A power-up
sequence will occur showing the unit type,
lighting up all the LEDs, showing the software
revision and finally displaying “IdL”.
3. Reset the carbon prefilter timer. (See “To Reset
Carbon Prefilter Timer.”)
4. Press the ʻStart/Stopʼ switch on the keypad to
start operation. Display will show “Add” and
the ʻRO System Operatingʼ LED will be illumi-
nated. The DI pump and UV lamp will not turn
on any time “Add” is displayed.
5. 2 Hour New Membrane Rinse-
Allow the unit to operate until the display
shows “Er7” (approximately 2 hours.) This will
rinse the new membrane of its preservatives. It
is possible that during this time the “Low RO
Purity” LED may illuminate.
6. After the membrane rinse the unit can be shut
down and the reservoir drain plug can be re-
inserted (insert completely then back out very
slightly).
Initial Operation
Note
Cartridge rinse up procedure must be
followed after each cartridge and/or fil-
ter replacement.
Note
For more demanding applications
where low TOC water is required, a
third reservoir volume rinse of the car-
tridges and filter may be necessary.
Warning
Use a properly grounded
electrical outlet of correct voltage
and current handling capacity.
Warning
This device is to be used with
potable water feeds only. Cleaning
agents must be used in compliance
with instructions in this manual.
Failure to comply with the above
could result in explosion and person-
al injury.
Note
The display will read “ADD” when
the reservoir is filling and water is
not running through DI cartridge.