unit to produce high quality water quickly upon being
placed in the Run Mode and prolong the life of the UV
lamp. It is recommended that the EASYpure R
O
D
I
be
placed in the Standby Mode during non-work hours. At
the end of the work day, press the STANDBY switch to
place the unit in Standby Mode. “SbY” will appear on the
display. Also in the Standby Mode, the RO operation auto-
matically turns on to fill the reservoir tank as needed.
Flush Mode
Manual Flush (RO Membrane)
Pressing the Flush switch once will initiate a one minute
flush of the RO membrane. During this one minute flush,
water is sent to drain through the drain tubing and the dis-
play will show FL1. Pressing the Flush switch twice will
initiate a five minute flush of the RO membrane. During
this five minute flush the display will count down FL5,
FL4, FL3, etc. After the timed flush has ended, the
EASYpure R
O
D
I
will return to the mode (“IdL”, Run or
“SbY”) it was in when you initiated the flush.
Automatic Flush (RO Membrane)
In the Run Mode, the EASYpure R
O
D
I
will initiate a one
minute automatic flush of the RO membrane once every
twenty-four hours of operation. This prevents buildup on
the RO membrane. A microcontroller timer in the
EASYpure R
O
D
I
counts the hours power is applied to the
unit. When the timer reaches twenty-four hours, the
EASYpure R
O
D
I
initiates a one minute flush if the unit is
in the Run and Standby Modes. If the unit was in
Standby, the automatic flush will occur immediately after
the START/STOP switch is pressed and Run Mode
entered. The timer will reset to zero when the one minute
automatic flush is completed.
Idle Mode (“IdL”)
“IdL” indicates the unit is powered and waiting to be
placed in Run, Standby or Flush Mode.
23
C
ONTROLS AND
N
ORMAL
O
PERATION
Note
Each time the RO system automati-
cally turns ON, in Run or Standby
mode, the RO membrane will be
flush water to the drain for 30 sec-
onds prior to filling the tank.