7
Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide
1.1 Methods
Notes:
1. For photometric determination of pH values only use BROMOCRESOLPURPLE tablets
in black printed foil pack and marked with METER.
2. pH values below 5.2 and above 6.8 can produce results inside the measuring range.
A plausibility test (pH-meter) is recommended.
3. The accuracy of the colorimetric determination of pH-values depends on various
boundary conditions (buffer capacity of the sample, salt contents etc.).
4. Salt error
Correction of test results (average values) for samples with salt contents of:
Indicator
Salt content
Bromcresolpurple
1 molar
– 0.26
2 molar
– 0.33
3 molar
– 0.31
The values of Parson and Douglas (1926) are based on the use of Clark and Lubs
buffers.
1 Mol NaCl = 58.4 g/l = 5.8 %
Содержание AQUAfast AQ3700
Страница 1: ...Thermo Scientific Orion AQUAfast AQ3700 Colorimetry Meter User Guide ...
Страница 8: ... Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Страница 9: ... Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 1 Methods ...
Страница 115: ...113 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide 1 1 Methods Note 1 Mn MnO4 KMnO4 ...
Страница 143: ...141 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Страница 193: ...191 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide 1 1 Methods Notes 1 SO3 Na2 SO3 ...
Страница 206: ...204 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Страница 207: ...205 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 2 Instrument Manual ...
Страница 210: ...208 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Страница 254: ...252 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Страница 255: ...253 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 3 Enclosure ...
Страница 263: ......