20
Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide
User Concentration Methods
It is possible to enter and store up to 10 User Concentration Methods.
Therefor you need 2 to 14 standards of known concentration and one blank (deionized water
or reagent blank value). The Standards should be measured with increasing concentrations
and from the brightest to the darkest coloration.
The measuring range for „Underrange“ and „Overrange“ is defined with –2600 mAbs* and
+2600 mAbs*. After selection of a method the concentration of the lowest and highest used
standard is displayed as measuring range. The operation range should be within this range
to achieve best results.
*1000 mAbs = 1 Abs = 1 E (displayed)
Entering a User Concentration:
Press [MODE], [Shift] + [6][4] keys.
Confirm with [ ] key.
Entry Procedure:
The display shows:
Enter a method number in the range from 850 to 859,
e.g.: [Shift] + [8][5][0]
Confirm with [ ] key.
Note:
if the entered number has already been used to save a
concentration the display shows the query:
• Press [Shift] + [0] or [ESC] key to go back to method no.
query.
• Press [Shift] + [1] key to start entry mode.
Enter the required wavelength, e.g.: [Shift] + [2] for
560 nm.
Press [ ] or [ ] keys to select the required unit.
Confirm with [ ] key.
< User concentr.>
choose no.: ____
(850-859)
choose unit:
>>
mg/l
g/l
mmol/l
mAbs
µg/l
E
A
%
Overwrite conc. meth.?
YES: 1, NO: 0
wavelength:
1: 530 nm
4: 430 nm
2: 560 nm
5: 580 nm
3: 610 nm
6: 660 nm
Mode
4
6
2
8
0
5
Содержание AQUAfast AQ3700
Страница 1: ...Thermo Scientific Orion AQUAfast AQ3700 Colorimetry Meter User Guide ...
Страница 8: ... Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Страница 9: ... Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 1 Methods ...
Страница 115: ...113 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide 1 1 Methods Note 1 Mn MnO4 KMnO4 ...
Страница 143: ...141 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Страница 193: ...191 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide 1 1 Methods Notes 1 SO3 Na2 SO3 ...
Страница 206: ...204 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Страница 207: ...205 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 2 Instrument Manual ...
Страница 210: ...208 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Страница 254: ...252 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Страница 255: ...253 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 3 Enclosure ...
Страница 263: ......