Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide
1.1 Methods
Notes:
1. Vial cleaning:
As many household cleaners (e.g. dishwasher detergent) contain reducing substances,
the subsequent determination of Chlorine dioxide may show lower results. To avoid
any measurement errors, only use glassware free of Chlorine demand.
Preparation: Put all applicable glassware into Sodium hypochlorite solution (0.1 g/l) for
one hour, then rinse all glassware thoroughly with deionized water.
2. Preparing the sample:
When preparing the sample, the escape of Chlorine dioxide gases, e.g. by pipetting or
shaking, must be avoided. The analysis must take place immediately after taking the
sample.
3. The DPD color development is carried out at a pH value of 6.2 to 6.5. The reagent
tablet therefore contains a buffer for the pH adjustment.
Strong alkaline or acidic water samples must be adjusted between pH 6 and pH 7
before the tablet is added (use 0.5 mol/l Sulfuric acid resp. 1 mol/l Sodium hydroxide).
4. Exceeding the measuring range:
Concentrations above 19 mg/l Chlorine dioxide can lead to results showing 0 mg/l.
In this event, the water sample must be diluted with water free of Chlorine dioxide.
10 ml of the diluted sample should be mixed with the reagent and the measurement
repeated.
5. If ??? is displayed at a differentiated test result see page 242.
Oxidizing agents such as Chlorine, Ozone etc. interfere as they react in the same way as
Chlorine dioxide.
Содержание AQUAfast AQ3700
Страница 1: ...Thermo Scientific Orion AQUAfast AQ3700 Colorimetry Meter User Guide ...
Страница 8: ... Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Страница 9: ... Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 1 Methods ...
Страница 115: ...113 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide 1 1 Methods Note 1 Mn MnO4 KMnO4 ...
Страница 143: ...141 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Страница 193: ...191 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide 1 1 Methods Notes 1 SO3 Na2 SO3 ...
Страница 206: ...204 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Страница 207: ...205 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 2 Instrument Manual ...
Страница 210: ...208 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Страница 254: ...252 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Страница 255: ...253 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 3 Enclosure ...
Страница 263: ......