1.1 Methods
Notes:
1. Strong alkaline or acidic water samples must be adjusted to approx. pH 7 before analysis
(use 1 mol/l Hydrochloric acid resp. 1 mol/l Sodium hydroxide).
2. If chlorine is known to be present, add one drop of 0.1 mol/l Sodium thiosulfate for each
0.3 mg/l Cl
2
in a one litre water sample.
3. Iron interferes with the test. The interferences will be eliminated as follows:
Determine the amount of total iron present in the water sample. Add an iron standard
solution with the same concentration to the vial (point 1) instead of deionized water to
produce the blank.
4. Conversion:
mg/l NH
4
= mg/l N x 1.29
mg/l NH
3
= mg/l N x 1.22
5. N
NH
4
NH
3
Содержание AQUAfast AQ3700
Страница 1: ...Thermo Scientific Orion AQUAfast AQ3700 Colorimetry Meter User Guide ...
Страница 8: ... Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Страница 9: ... Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 1 Methods ...
Страница 115: ...113 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide 1 1 Methods Note 1 Mn MnO4 KMnO4 ...
Страница 143: ...141 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Страница 193: ...191 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide 1 1 Methods Notes 1 SO3 Na2 SO3 ...
Страница 206: ...204 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Страница 207: ...205 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 2 Instrument Manual ...
Страница 210: ...208 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Страница 254: ...252 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Страница 255: ...253 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 3 Enclosure ...
Страница 263: ......