23
Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide
Delete user calibration
This chapter only applies for methods which can be user calibrated.
Select the required method.
Instead of zeroing the instrument press [MODE], [Shift] +
[4][6] keys and confirm with [ ] key.
The display shows:
Press [Shift] + [1] key to delete user calibration.
Press [Shift] + [0] key to keep the valid user calibration.
The instrument goes back to Zero-query automatically.
<user calibration>
100
Chlorine T
0.02-6 mg/l Cl2
clear user
calibration?
YES: 1, NO: 0
100
Chlorine T
0.02-6 mg/l Cl2
prepare ZERO
press ZERO
Mode
6
4
1
0
Содержание AQUAfast AQ3700
Страница 1: ...Thermo Scientific Orion AQUAfast AQ3700 Colorimetry Meter User Guide ...
Страница 8: ... Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Страница 9: ... Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 1 Methods ...
Страница 115: ...113 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide 1 1 Methods Note 1 Mn MnO4 KMnO4 ...
Страница 143: ...141 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Страница 193: ...191 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide 1 1 Methods Notes 1 SO3 Na2 SO3 ...
Страница 206: ...204 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Страница 207: ...205 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 2 Instrument Manual ...
Страница 210: ...208 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Страница 254: ...252 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Страница 255: ...253 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 3 Enclosure ...
Страница 263: ......