Antaris EX-Z1 Specifications
Utilities
Each system requires an AC electrical service, fiber optic cabling for sample
data collection, Ethernet for remote computer and spectrometer control,
and nitrogen or air for purge and cooling. Either gas must be cleaned and
dried. Depending on your application, optional analog and digital PLC
transmission wiring also may be used. All wire must comply with local
codes.
Item
Specification
Electrical
AC service
3-wire: live, neutral, earth ground
Dedicated, switched, 15A circuit
Material
10–14 AWG (2–5 mm
2
) copper wire
Tighten input terminals to 0.5 Nm torque
Input
voltage
200 to 264 VAC
Input current
3.5 A max
Startup current
15 A max
Line frequency
50 to 60 Hz
Line disturbance
Sags, surges or other line disturbances
must not exceed 10% of input voltage
(even for a half cycle)
Noise
< 2 volts (common mode)
< 20 volts (normal mode)
Fiber optic
Connections
SMA905
Bend radius
per manufacturer’s specification
Continued…
80
Antaris EX Safety, Installation, and Site Preparation Guide
Thermo Fisher Scientific
Содержание Antaris EX
Страница 1: ...Antaris EX Safety Installation and Site Preparation Guide ...
Страница 6: ...This page intentionally left blank ...
Страница 10: ...This page intentionally left blank ...
Страница 29: ...Safety Considerations Thermo Fisher Scientific Antaris EX Safety Installation and Site Preparation Guide 23 ...
Страница 30: ...Safety Considerations OR 24 Antaris EX Safety Installation and Site Preparation Guide Thermo Fisher Scientific ...
Страница 32: ...Safety Considerations 26 Antaris EX Safety Installation and Site Preparation Guide Thermo Fisher Scientific ...
Страница 35: ...Safety Considerations s Thermo Fisher Scientific Antaris EX Safety Installation and Site Preparation Guide 29 ...
Страница 36: ...Safety Considerations s 30 Antaris EX Safety Installation and Site Preparation Guide Thermo Fisher Scientific ...
Страница 92: ...This page intentionally left blank ...
Страница 108: ...269 188702 Rev A ...