Installation Process
3.
Use a 4 mm OD nylon tube to connect a temporary pressure gauge
or water manometer to the system pressure test point.
Press down on the quick connect collar and insert nylon tubing into
the test point. Release the collar to secure the tubing.
4.
Use a flat-blade screwdriver to open the leakage compensation valve
completely.
Turn the valve counterclockwise as far as possible.
Leakage
compensation
valve
5.
Slowly open the supply regulator and allow the enclosure pressure
to rise until the relief valve opens.
You will hear gas flowing through the relief valve once the valve opens.
Note
The relief valve contains a spark arrestor.
6.
Check the pressure gauge or water manometer to be sure that the
relief valve opened at or below the maximum pressure indicated on
the system purge label.
Thermo Fisher Scientific
Antaris EX Safety, Installation, and Site Preparation Guide
57
Содержание Antaris EX
Страница 1: ...Antaris EX Safety Installation and Site Preparation Guide ...
Страница 6: ...This page intentionally left blank ...
Страница 10: ...This page intentionally left blank ...
Страница 29: ...Safety Considerations Thermo Fisher Scientific Antaris EX Safety Installation and Site Preparation Guide 23 ...
Страница 30: ...Safety Considerations OR 24 Antaris EX Safety Installation and Site Preparation Guide Thermo Fisher Scientific ...
Страница 32: ...Safety Considerations 26 Antaris EX Safety Installation and Site Preparation Guide Thermo Fisher Scientific ...
Страница 35: ...Safety Considerations s Thermo Fisher Scientific Antaris EX Safety Installation and Site Preparation Guide 29 ...
Страница 36: ...Safety Considerations s 30 Antaris EX Safety Installation and Site Preparation Guide Thermo Fisher Scientific ...
Страница 92: ...This page intentionally left blank ...
Страница 108: ...269 188702 Rev A ...