Installation Process
You may make additional modifications to the gland plates to suit your
installation. See the schedule drawings and schematics that accompanied
the enclosure for details.
AC power should be provided on a switched outlet to allow the power-up
sequence to be completed conveniently.
Wash down,
hose down applications
As shipped from the factory, the instrument AC power cord provided does
not provide any water protection at the plug end. You may use either a
waterproof outlet box or remove the rewirable plug from the cord for direct
wiring applications. The PLC I/O signals must be connected through cable
glands that maintain the rating of the enclosure. These ports are located on
the top of the enclosure. You may make additional modifications to the
gland plates provided to suit your installation. See the schedule drawings
later in this manual for locations and specifications.
Sealed USB and Ethernet connectors are provided. Unsealed cables will
connect to these components. Sealed mating connectors and cables are
available from the Bulgin Components Buccaneer line of products.
If the integrated computer option is installed, there will be 2 USB
connectors available for peripheral devices such as printers, drives, and
scanners.
There will be 2 Ethernet connectors. If the remote computer option is
installed, the “computer” connector will be for connection to the
spectrometer. If the integrated computer option is installed, the
“computer” connector connects to the computer. If the PLC option is
installed, the “PLC” connector connects to the PLC.
Internal enclosure connections
Inside the enclosure, connect the input/output wires according to your
particular options. See the schedule drawings and schematics for the
analyzer internal wiring for specific details. All connections inside the
enclosure should be made to the DIN rail blocks as shown in the schedule
drawings.
Thermo Fisher Scientific
Antaris EX Safety, Installation, and Site Preparation Guide
39
Содержание Antaris EX
Страница 1: ...Antaris EX Safety Installation and Site Preparation Guide ...
Страница 6: ...This page intentionally left blank ...
Страница 10: ...This page intentionally left blank ...
Страница 29: ...Safety Considerations Thermo Fisher Scientific Antaris EX Safety Installation and Site Preparation Guide 23 ...
Страница 30: ...Safety Considerations OR 24 Antaris EX Safety Installation and Site Preparation Guide Thermo Fisher Scientific ...
Страница 32: ...Safety Considerations 26 Antaris EX Safety Installation and Site Preparation Guide Thermo Fisher Scientific ...
Страница 35: ...Safety Considerations s Thermo Fisher Scientific Antaris EX Safety Installation and Site Preparation Guide 29 ...
Страница 36: ...Safety Considerations s 30 Antaris EX Safety Installation and Site Preparation Guide Thermo Fisher Scientific ...
Страница 92: ...This page intentionally left blank ...
Страница 108: ...269 188702 Rev A ...