Installation Process
Check the label on the front of the analyzer if you are unsure of the
correct normal operating pressure.
2.
Check the alarm/pressurized and purging indicators.
As the pressure builds, the alarm/pressurized indicator should change
from red to green. The purging indicator should change from black to
yellow whenever purge gas flows into the system.
Purging (Yellow)
Alarm/pressurized (Green)
3.
Turn off the purge supply at its source and wait for the
Alarm/Pressurized indicator to turn red and the Purging indicator
to turn black.
4.
Repeat steps 1 through 3 several times to confirm indicator
operation.
Thermo Fisher Scientific
Antaris EX Safety, Installation, and Site Preparation Guide
45