Installation Process
5.
Move the lever to the Purge Off position.
6.
Apply power to the branch circuit.
Antaris EX-IP
In general, the power on procedure for EX-IP installations includes these
steps:
Danger
Avoid explosion and fire hazards. EX-IP systems offer no hazardous
location protection. Never apply power to an EX-IP system in a volatile
atmosphere.
Warning
Avoid shock hazard. Always use a power outlet that maintains the hose
down rating of the analyzer.
1.
Connect the power cord to the EX-IP rated, switched outlet that
services the instrument.
2.
Apply power to the outlet.
Powering off
Specific information about power-off procedures, as well as component
cooling cycles and times, is included in documentation that accompanied
the enclosure and on labels on the enclosure itself. Use these instructions
whenever you wish to power off the instrument.
Danger
Avoid explosion and fire hazard. The NIR source is very hot. In volatile
environments always wait at least the amount of time specified on the
labeling on the enclosure before opening the enclosure.
This allows internal components to cool to a safe level.
70
Antaris EX Safety, Installation, and Site Preparation Guide
Thermo Fisher Scientific
Содержание Antaris EX
Страница 1: ...Antaris EX Safety Installation and Site Preparation Guide ...
Страница 6: ...This page intentionally left blank ...
Страница 10: ...This page intentionally left blank ...
Страница 29: ...Safety Considerations Thermo Fisher Scientific Antaris EX Safety Installation and Site Preparation Guide 23 ...
Страница 30: ...Safety Considerations OR 24 Antaris EX Safety Installation and Site Preparation Guide Thermo Fisher Scientific ...
Страница 32: ...Safety Considerations 26 Antaris EX Safety Installation and Site Preparation Guide Thermo Fisher Scientific ...
Страница 35: ...Safety Considerations s Thermo Fisher Scientific Antaris EX Safety Installation and Site Preparation Guide 29 ...
Страница 36: ...Safety Considerations s 30 Antaris EX Safety Installation and Site Preparation Guide Thermo Fisher Scientific ...
Страница 92: ...This page intentionally left blank ...
Страница 108: ...269 188702 Rev A ...