Installation Process
Commissioning
an Antaris EX-Z2
Before bringing the analyzer on line for the first time, you must clear the
supply line of debris and condensation, verify that the pressure indicators
and alarms are operating correctly, and in some cases, fine-tune the leakage
compensation valve. You will need the following tools for these procedures:
•
flat-blade screwdriver
Clearing the purge supply line
To remove debris and condensation from the purge gas supply line:
1.
Disconnect the supply line from the inlet to the purge system.
2.
Blow clean, dried air or nitrogen through the supply line for at least
15 seconds per meter (3 feet) of supply pipe length.
This removes any debris or condensation.
3.
If there are questions about the supply line pressure, connect a
pressure gauge and confirm that the supply line pressure being
delivered to the enclosure inlet is between 4 barg and 8 barg, 60 psi
and 115 psi, or 400 kPA and 800 kPA.
4.
Turn off the purge gas and reconnect the supply line to the purge
inlet.
Confirming alarm
and indicator operation
Each Antaris EX-Z2 analyzer includes a purge system and pressure sensors.
To verify that the system is operating properly:
1.
Turn on the purge gas supply regulator to normal operating
pressure.
Thermo Fisher Scientific
Antaris EX Safety, Installation, and Site Preparation Guide
49
Содержание Antaris EX
Страница 1: ...Antaris EX Safety Installation and Site Preparation Guide ...
Страница 6: ...This page intentionally left blank ...
Страница 10: ...This page intentionally left blank ...
Страница 29: ...Safety Considerations Thermo Fisher Scientific Antaris EX Safety Installation and Site Preparation Guide 23 ...
Страница 30: ...Safety Considerations OR 24 Antaris EX Safety Installation and Site Preparation Guide Thermo Fisher Scientific ...
Страница 32: ...Safety Considerations 26 Antaris EX Safety Installation and Site Preparation Guide Thermo Fisher Scientific ...
Страница 35: ...Safety Considerations s Thermo Fisher Scientific Antaris EX Safety Installation and Site Preparation Guide 29 ...
Страница 36: ...Safety Considerations s 30 Antaris EX Safety Installation and Site Preparation Guide Thermo Fisher Scientific ...
Страница 92: ...This page intentionally left blank ...
Страница 108: ...269 188702 Rev A ...