3
Installation
Installation of the AI 3000/AS 3000 II on the GC
Thermo Scientific
AI 3000/AS 3000 Sampling System - Operating Manual
51
Installation of the Syringe
To install the syringe follow the instructions reported below.
The installation of the syringe is a simple operation. However, it must be performed with
caution to avoid damages to the syringe needle and ensure an optimal performance of the
injection device.
The standard syringe has 10
μ
L capacity with 50 mm needle length. 5 and 0.5
μ
L optional
syringes are optionally available.
1. Open the safety door of the turret.
2. Insert the syringe needle into the vial capture device. See
Figure 46.
Syringe Installation (1)
3. Accommodate the syringe body into its seat paying attention to insert simultaneously the
flange and head of the syringe plunger into their relevant guides.
4. Lock the syringe by turning clockwise the lock knob by approximately 180°. See
5. Close the safety door of the rotating turret.
1
2
3
4
5
6
7
8
9
1 = Syringe Needle
2 = Syringe Body
3 = Flange
4 = Plunger
5 = Plunger Head
6 = Vial Capture Device
7 = Syringe Body Seat
8 = Flange Guide
9 = Plunger Head Guide