Installation
Installing the Sampling System Support on the TRACE 1300/TRACE 1310
34
AI 3000/AS 3000 Sampling System - Operating Manual
Thermo Scientific
Assembling of the Sampling System Support on the Back Injector Module
The following is the description of how to mount the sampling system support on the
back
injector module of the TRACE 1300/TRACE 1310. See
.
Figure 31.
Assembling on the Back Injector Module (1)
1. Remove from the GC top cover the three plastic caps covering the corresponding fixing
holes. See A, B, and C.
2. Orient the L-shape bottom plate to align the three installation holes to the corresponding
fixing holes located on the GC top.
3. Insert the provided fixing screws into the holes present on the L-shape bottom plate.
Note
In comparison with the installation on the front injector module, the
installation on the back injector module requires to upturn the L-shape bottom plate
before fixing it on the GC.
A
B
C
L-shape Bottom Plate