4
AI 3000/AS 3000 II Control
Guidelines for Programming with Different Injectors
74
AI 3000/AS 3000 Sampling System - Operating Manual
Thermo Scientific
PTV Inlet with a Merlin Microseal™
The above PTV conditions apply also when the Merlin Microseal™ is used.
PKD and PPKD Inlets
For injection into packed column the standard insertion depth with 3-5 seconds of Pre-dwell
time is suggested. See
Note
In order to guarantee long-life performance of the Merlin Microseal™ it is
recommended to use syringes with 0.63 mm needle diameter and truncated cone shape on
tip. Please refer to the spare part catalog for part number.
Table 17.
PKD and PPKD Inlets
Injector
AI 3000/AS 3000 II Injection Parameters
Packed Injector (PKD)
Purged Packed Injector (PPKD)
Standard Injection Depth
3-5 sec. Pre-Dwell Time
0 sec. Post-Dwell Time