4
AI 3000/AS 3000 II Control
Control Through the TRACE GC Ultra
64
AI 3000/AS 3000 Sampling System - Operating Manual
Thermo Scientific
Extended Control Menu
In this menu it is possible to set how long the syringe needle must remain inside the injector
without injecting. This allows to avoid discriminations caused by evaporation of the sample
contained in the syringe needle. In this way the sample is completely drawn into the syringe
and the needle left inside the injector and heated for a few seconds before injecting. See
for details.
Table 9.
Control Through the TRACE GC Ultra: Extended Control Menu
Menu
Range
Comments
EXTENDED
CONTROL
This line shows the menu title.
Viscous sample
Yes, No
This parameter defines the speed at which the
sample is drawn from the vial as a function of
the sample viscosity. Select No (default value) if
the sample has low viscosity. Select Yes if the
sample has high viscosity.
Sampl. depth
Bottom,
Center
selectable in
the submenu
This parameter determines the penetration
depth of the syringe needle into the vial.
Press
Mode/Type
or
Enter
to enter the
submenu.
• Selecting
Bottom
(default value) the needle
goes down to the vial bottom.
• Selecting
Center
the needle goes down to
half vial.
Inj. Depth
Standard,
Minimum
selectable in
the submenu
This parameter determines the penetration
depth of the syringe needle into the injector.
Press
Mode/Type
or
Enter
to enter the
submenu.
• Selecting
Standard
(default value) the
needle goes down to the maximum depth
allowed.
• Selecting
Minimum
the needle enters the
injector and stops immediately beyond the
septum (Cold Needle Technique).