2
Sampling Unit
Sample Trays
16
AI 3000/AS 3000 Sampling System - Operating Manual
Thermo Scientific
• 15-pin female connector (double density) marked
TRAY
for the communication between
the sampling unit and the motor actuating the hub positioned on the supporting plate of
the 105-position sample tray.
• 9-pin female connector marked
RS232
for the communication between the
autoinjector/sampler and the GC (connector marked
AUTOSAMPLER
) through a
proper cable.
• 6-pin female connector marked
GC
for connecting the autoinjector/sampler to the GC
(connector marked
SAMPLER SIGNAL
) through the proper cable.
• 4-pin male connector marked
TWIN SYNC
for the synchronism between two
AI 3000/AS 3000 II installed in Gemini configuration through the proper cable.
Sample Trays
The AI 3000/AS 3000 II can accept two types of sample trays which are accommodated in
one of the two side housings on the base.
The type of sample tray accessory installed on the instrument (8 or 105 positions) is
automatically acknowledged by the autoinjector/sampler when switched on, and it determines
the instrument configuration.
•
AI 3000
uses the 8-position sample tray.
•
AS 3000
uses the 105-position sample tray.
8-Position Sample Tray
The 8-position tray, see
, consists of a fixed arc-shaped support and can contain up
to 8 vials numbered
1
to
8
. The slot present on the tray determines the position
1
and allows
to inspect the level of the liquid contained in the vial.
Figure 14.
8-position Sample Tray
105-Position Sample Tray Assemble
The assemble, see
, consists of a dedicated supporting plate provided with a hub and
a 105-position removable sample tray.