3
Installation
Installing the Sampling System Support on the FOCUS GC
48
AI 3000/AS 3000 Sampling System - Operating Manual
Thermo Scientific
Figure 41.
Two or More FOCUS Placed on the Same Workbench
Proceed as follows:
1. Remove from the GC upper cover the two plastic caps covering the corresponding fixing
holes.
2. Insert into each slot present on the support the provided fixing screw.
3. Guide the two fixing screws into the corresponding fixing holes located on the GC upper
panel.
4. Loosely tighten the screws.
A
B
1
2
3
2
1
1
3
4
4
A = Conventional Installation
B Space Saving Installation
1 = Fixing Screws
2 = Sampling System Support
3 = AS 3000
4 = Second FOCUS GC
Содержание AI 3000