Sampling Unit
Sample Trays
Thermo Scientific
AI 3000/AS 3000 Sampling System - Operating Manual
17
The rotation of the tray is accomplished by a motor, located under the supporting plate,
which actuates the hub thanks to suitable commands coming from the sampling unit via an
interfacing cable.
An optical sensor defines the rotation travel end and the vial position.
Figure 15.
Supporting Plate and 105-Position Sample Tray
The tray can contain up to 105 vials numbered
01
to
105
arranged on three circular rows.
• Vials
01
to
40
are placed into the external row.
• Vials
41
to
80
are placed into the middle row.
• Vials
81
to
105
are placed into the internal row.
Vials
Vials of 0.3 and 2 mL capacity can be used.
The 0.3 mL vial is exactly the same as the 2 mL vial, provided inside with a 0.3 mL calibrated
insert.
Supporting Plate
Hub
105-position Sample Tray
CAUTION
Vials must be accurately closed using appropriate septa and ring nuts.
Septa must be those recommended by Thermo Fisher Scientific. The use of septa with
different characteristics might damage or bend the syringe needle.
During the vials preparation, it is recommended to comply with applicable safety
regulations, specially as far as the conditions of the workplace ventilation are concerned.