Installing the Sampling System Support on the FOCUS GC
46
AI 3000/AS 3000 Sampling System - Operating Manual
Thermo Scientific
Installing the Sampling System Support on the FOCUS GC
To install the support on the FOCUS GC, please refer to the instructions reported below.
• See
Assembling of the Sampling System Support on the GC right side
• See
Assembling of the Sampling System Support on FOCUS GC MS Version
• See
Assembling of the Sampling System Support on the GC right side
The following is the description of how to mount the sampling system support in the default
configuration, i.e. on the
right
side of the FOCUS GC.
1. Remove from the GC upper cover the two plastic caps covering the corresponding fixing
holes.
2. Insert into each slot present on the support the provided fixing screw.
3. Guide the two fixing screws into the corresponding fixing holes located on the GC upper
panel.
4. Loosely tighten the screws.
Assembling of the Sampling System Support on FOCUS GC MS Version
The installation on the GC
left
side, instead the conventional installation on the GC right
side, may be carried out to reduce the working area of your FOCUS GC MS Version. Refer to
the
Figure 38.
Sampling System Support on the FOCUS GC MS Version (1)
Proceed as follows:
A
1
2
1
B
1
4
3
A = Conventional Installation
B = Left Side Installation
1 = Right Side Fixing Hole
2 = Support Positioning
3 = Left Side Fixing Holes
4 = Support Positioning