3.
Wash the bead removal bag with wash buffer to transfer the remaining cells in the bead removal
bag to the output bag.
4.
Add air to the bead removal bag to push the remaining liquid and cells in the bead removal bag to
the output bag.
Remove Single-Use Kit
CAUTION!
Seal all fluid lines using a tube sealer or manual clamps prior to opening the pinch valves
to prevent fluid movement when the instrument pinch valves are released.
1.
Seal off finished product from the Single-Use Kit.
2.
From the home screen select
Open valves
. Remove the tubes from the pinch valves.
Chapter 6
Basic instrument operation
Remove Single-Use Kit
6
152
CTS
™
DynaCellect
™
Magnetic Separation System User Guide