4.
Take all tubes out of the catch on the top of the cassette.
CAUTION!
The CTS
™
DynaCellect
™
Single-Use Kit (Cat. Nos.
and
) have been
validated for one-time use by customers. Single-Use Kits are not recommended to be washed,
re-sterilized, or reused as sterility and quality of these kits have not been validated for multiple uses.
Connecting vessels, bags and reagents
The instrument has been specifically designed for single-use, functionally closed processing. The
CTS
™
DynaCellect
™
Cell Isolation Kit and CTS
™
DynaCellect
™
Bead Removal Kit use DEHP-Free PVC,
specifically Tygon ND-100-65 size of 3/32" I.D. x 5/32" O.D. x 1/32" Wall to enable sterile welding to
compatible bags using devices that service the blood bank industry such as the TSCD II Sterile Tubing
Welder (Terumo BCT). See the instructions provided by instrument providers for completion of sterile
welding.
Chapter 6
Basic instrument operation
Connecting vessels, bags and reagents
6
CTS
™
DynaCellect
™
Magnetic Separation System User Guide
127