(continued)
Code
Message
Context
Description
Recovery action
0x8871
Emergency stop is
activated
Idle
Recipe
Manual
Emergency stop
was activated
Reset emergency
stop switch and
perform recovery
using the reset
button in the
user interface
prompt dialog after
releasing the switch
0x8810
Rocker lid is not
closed
Recipe
Rocker bag lid is
not closed when
starting a run
Export service logs
and send to tech
support for further
assistance
0x8811
No kit detected
Recipe
No kit detected
when a run when
starting a run
Ensure that the
tube is stretched
before inserting into
pressure sensors
0x8812
Pressure sensor is
out of range
Recipe
Occlusion pressure
sensor is out of
detection range
Ensure that the
tube is stretched
before inserting into
pressure sensors
0x8813
Rocker bed leak
detected
Recipe
Leak sensor in
rocker nest detects
a leak
Check for leaks in
rocker nest and
clean the leak
accordingly
0x8814
Pump is not closed
Recipe
Pump cover is
not closed when
starting a run
Ensure that the
pump cover is
closed before
starting a run
0x8815
Tube occlusion
detected
Recipe
• Pressure
sensor
exceeded
threshold and
detects a
possible tube
occlusion event
while a recipe
run is in
progress
• Recipe run is
aborted
Investigate for tube
blockage in fluidics
path
Appendix B
Error lists
System error list
B
164
CTS
™
DynaCellect
™
Magnetic Separation System User Guide
Содержание A55867