5.
In the
Isolation Protocol Kit: Details
screen, the sites can be labelled (the site label does not
influence the instrument behavior; it only informs users what content should be on the specific
lines for the protocol in question. Site A Isolation Bag is default when the isolation kit is chosen).
Select
Bead location
to proceed.
6.
In the
Select All Bead Site Locations
screen, select site location of your beads (the identification
of the bead site is needed to update the batch-specific bead volume in
Run Protocol
mode as
output from the bead calculator). Select
Edit volumes
to proceed.
Chapter 3
Embedded graphical user interface (eGUI) user instructions
Manage protocols
3
CTS
™
DynaCellect
™
Magnetic Separation System User Guide
51