Operation
Calibration
3-24
42iQTL Instruction Manual
Thermo Scientific
The Adjust NO, NO
x
, and Prereactor Background screens are used to
adjust the instrument zero background. Before making an adjustment, be
sure the analyzer samples zero air for at least 5 minutes.
It is important to note the averaging time when calibrating. The longer the
averaging time the more precise the calibration results. To achieve
maximum precision, allow the instrument to stabilize each time input gas is
changed and set the averaging time to 300-second averaging.
Home Screen>Calibration>Advanced Calibration>Manual
Calibration>Adjust Backgrounds>Adjust NO Background
Home Screen>Calibration>Advanced Calibration>Manual
Calibration>Adjust Backgrounds>Adjust NO
x
Background
Adjust NO, NO
x
and
Prereactor Background
Содержание 42iQTL
Страница 1: ...42iQTL Instruction Manual Trace Level NO NO2 NOx Analyzer 119019 00 1April2020 ...
Страница 2: ......
Страница 26: ......
Страница 34: ...Operation Instrument Display 3 8 42iQTL Instruction Manual Thermo Scientific ...
Страница 162: ......
Страница 190: ......
Страница 196: ...Maintenance Pump Rebuilding 5 6 42iQTL Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 212: ...Servicing Fan Replacement 7 8 42iQTL Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 229: ...Servicing Power Supply Replacement Thermo Scientific 42iQTL Instruction Manual 7 25 Figure 7 25 Replacing Power Supply ...
Страница 253: ...Servicing DMC Ozonator Thermo Scientific 42iQTL Instruction Manual 7 49 Figure 7 49 Remove Ammonia Scrubber ...
Страница 258: ......
Страница 292: ......
Страница 304: ......
Страница 308: ......
Страница 309: ......